Specifications
System Overview
- Data Storage
- 30MB internal memory; up to 5,000 spectra
- Training Requirements
- < 10 mins for operator; 1 day for advanced user
Environmental
- Operating Humidity
- 93% RH @ 35 °C (non-condensing)
- Operating Temperature
- -4 to 122 °F (-20 to 50 °C)
- Storage Temperature
- 14 to 95 °F (-10 to 35 °C)
Physical Features
- Dimensions (L x W x H)
- 5.7 x 2.2 x 1.9 in (14.5 x 5.6 x 4.8 cm)
- Weight
- ≤0.88 lb (0.4 kg)
- Enclosure & Protection
- Plastic injection with rubber overmold; protection rating IP67 according to IEC 60529
Power
- Battery Specs
- Internal single cell Li-ion; operational battery life ≥36h; replaceable back-up CR123A with ≥18h operational battery life; recharge time ≤6h when using AC or USB power source
- Input Voltage
- 100-240 VAC (wall and car adapters and USB cable supplied)
- Start Up Time
- < 2 mins from cold start
Sampling & Analysis
- Sample Introduction
- Absorption of EM gamma emissions
- Sampling & Analysis
- From a few seconds to minutes
- Standards Compliance
- ANSI N42.32 PRD standard fully compliant, ANSI N42.48 SPRD standard fully compliant, including nuclide identification
- Threats
- Detects gamma radiation emitted from natural occurrences in the environment, special nuclear material, industrial, or medical material
System Interface
- Communication
- USB 2.0; mini-B socket; Bluetooth® ≤10m range (removable)
- Software
- On-board webserver software
- Display and Alerts
- Memory LCD black and white display
Technology
- Dose Rate & Accuracy (Cs-137)
- ≤100 nSv/h – 250 µSv/h (≤10µRem/h – 25 mRem/h); ±20%
- Energy Range [Gamma]
- 25 keV - 3 MeV
- Gamma [CsI]
- (18mm)³ Cesium Iodide (CsI) with Silicon Photomultiplier (SiPM)
- Gamma Spectrum
- 1,024 channels; 3 MeV
- Service Interval
- Recommended five year factory maintenance interval; annual dose rate calibration interval
- Technology
- Spectroscopic personal radiation detector (SPRD)
- Typical Resolution
- ≤7.5 % FWHM at 662 keV