Support Center
identiFINDER R100
For Support
Please
call (877) 692-2120
or
click here for support
For Service
Submit RMA
Resources:
- identiFINDER R100 Datasheet
- Download
Specifications
Environmental
- Operating Humidity
- 93% RH @ 35 °C non-condensing
- Operating Temperature
- -4 to 122 °F (-20 to 50 °C)
- Storage Temperature
- 14 to 95 °F (-10 to 35 °C)
Physical Features
- Dimensions (L x W x H)
- 5.3 × 2.2 × 1.5 in (13.5 × 5.5 × 3.7 cm)
- Weight
- 0.55 lb (0.25 kg)
- Enclosure & Protection
- Plastic with rubber overmold; protection rating IP67 according to IEC 60529
Power
- Battery Specs
- Two replaceable 1.5V AA Alkaline batteries; operational life >125 hours
- Input Voltage
- 4.5V to 5.5V USB micro-AB socket
- Start Up Time
- < 2 mins from cold start
Sampling & Analysis
- Compliance
- Meets the 1.5 m drop criteria required by ANSI N42.32, fully compliant with ANSI N42.33
- Sample Introduction
- Absorption of EM gamma emissions
- Sampling & Analysis
- Continuous
- Threats
- Detects gamma radiation emitted from natural occurrences in the environment, special nuclear material, industrial, or medical material
System Interface
- Alerts
- Visible (front and top LEDs), audible, and haptic (vibration)
- Communication
- USB 2.0; micro-AB socket; integrated Bluetooth® up to 10 m range
- Display
- Transflective Monochrome LCD
- Localization
- Language translation available worldwide
- Software
- On-board webserver software
Technology
- Dose Rate & Accuracy (Cs-137)
- ±20%; 50 µrem/h to 1 rem/h
- Dose Rate Range
- 5.0µrem/h (0.05µSv/h) to 1 rem/h (10mSv/h)
- Dose Rate Units
- rem/h or Sv/h; user adjustable
- Energy Range [Gamma]
- 25keV - 3MeV; energy compensated dose rate 60keV - 1.5MeV
- Gamma Sensitivity
- 17cps / mR/h
- Gemma (G-M)
- Energy Compensated Geiger-Mller Tube
- Service Interval
- Recommended five year factory maintenance interval; annual dose rate calibration interval
- Technology
- Personal Survey Meter (PSM)