Support Center
FLIR A315 f
For Support
Please
click here for support
For Service
Call 0049 69-950090 0
or
Submit RMA
Overview:
Trouble Shooting
I have a FLIR A3xx camera but I cannot see the image stream. What is wrong?
Read the StoryRegistration
I'm trying to register my FLIR product but I recieve an error that the serial number is not found. What is the reason?
Read the StoryConfiguration
My Ethernet camera (FLIR A3x0/A3x5/A6x5/Ax5) in FLIR Tools does not display a live image but instead shows a rainbow-colored image. How do I fix this?
Read the StoryInstructions
Does FLIR offer an SDK to control and get video stream from the FLIR IR cameras?
Learn moreTrouble Shooting
I cannot connect to my FLIR A3xx camera via Ethernet. What is the reason?
Read the StoryInstructions
I see that there are two firmware versions for the A3xx cameras. Which one should I choose?
Read the StorySystem Requirements
At what speed do we measure with the A-series cameras? ex. How fast could I detect a change in temperature and set a digital alarm?
Learn moreResources:
- Warranty: FLIR International Uncooled Thermal Camera 2-10 Limited Warranty
- Download
Specifications
Overview
- Automatic Heater
- Clears window from ice
- Base Mounting
- TBA
- Detector Type
- Focal Plane Array (FPA), uncooled microbolometer
- Housing material
- Aluminum
- IR Resolution
- 320 × 240 pixels
- Packaging Size
- 534 x 207 x 230 mm (21.0 x 8.1 x 9.1 in.)
- Size (L x W x H)
- 460 × 140 × 159 mm (18.1 × 5.5 × 6.3 in.)
- Spectral Range
- 7.5–13 µm
Connections & Communications
- Digital I/O connector type
- 6-pole jackable screw terminal
- Digital I/O isolation voltage
- 500 VRMS
- Digital I/O supply voltage
- 12/24 VDC, max 200 mA
- Digital input purpose
- Image tag (start, stop, general), Image flow control, (stream on/off), Input ext. device (programmatically read)
- Digital Inputs
- 2 opto-isolated, 10–30 VDC
- Digital Output Purpose
- Output to ext. device (programmatically set)
- Digital Outputs
- 2 opto-isolated, 10–30 VDC, max 100 mA
- Ethernet
- Control and image
- Ethernet Communication
- TCP/IP socket-based FLIR proprietary and GenICam protocol
- Ethernet Connector Type
- RJ-45
- Ethernet Image Streaming
- 16-bit 320 × 240 pixels @ 60 Hz - Signal linear - Temperature linear - Radiometric GigE Vision and GenICam compatible
- Ethernet Protocols
- TCP, UDP, SNTP, RTSP, RTP, HTTP, ICMP, IGMP, ftp, SMTP, SMB (CIFS), DHCP, MDNS (Bonjour), uPnP
- Ethernet Standard
- IEEE 802.3
- Ethernet Type
- Gigabit Ethernet
- Ethernet Video Streaming
- MPEG-4, ISO/IEC 14496-1 MPEG-4 ASP@L5
Environmental
- Atmospheric transmission correction
- Automatic, based on inputs for distance, atmospheric temperature and relative humidity
- Bump
- 5 g, 11 ms (IEC 60068-2-27)
- EMC
- EN 61000-6-2 (Immunity), EN 61000-6-3 (Emission), FCC 47 CFR Part 15 Class B (Emission)
- Emissivity Correction
- Variable from 0.01 to 1.0
- Encapsulation
- IP 66 (IEC 60529)
- Humidity (Operating and Storage)
- IEC 60068-2-30/24 h 95% relative humidity +25°C to +40°C (+77°F to +104°F)
- Operating Temperature Range
- –25°C to +50°C (–13°F to +122°F)
- Storage Temperature Range
- –40°C to +70°C (–40°F to +158°F)
- Vibration
- 2 g (IEC 60068-2-6)
Imaging & Optical
- Detector Pitch
- 25 µm
- Detector Time Constant
- Typical 12 ms
- f-number
- 1.3
- Focus
- Automatic or manual (built in motor)
- Image Frequency
- 30 Hz
- Lens Identification
- Automatic
Measurement & Analysis
- Accuracy
- ±4°C (±7.2°F) or ±4% of reading
- External Optics & Windows Correction
- Automatic, based on input of optics/window transmission and temperature
- Measurement Corrections
- Global object parameters
- Object Temperature Range
- –20 to +120°C (–4 to +248°F) 0 to +350°C (+32 to +662°F)
- Thermal Sensitivity/NETD
- < 0.05°C @ +30°C (+86°F) / 50 mK
Power
- External Power Connector Type
- 2-pole jackable screw terminal
- External Power Connector Type [heater]
- 2-pole jackable screw terminal
- External Power Operation
- The camera operates on 12/24 VDC, 9 W max. (allowed range: 10-30 VDC) and heaters on 24 VDC, 25 W max. In total: 34 W.
- External Power Operation [heater]
- 24 VDC, 25 W max.
- Voltage
- Allowed range 10–30 VDC
- Voltage [heater]
- Allowed range 21-30 VDC