Specifications
System Overview
Data Storage
30MB internal memory; up to 5,000 spectra
Training Requirements
< 10 mins for operator; 1 day for advanced user
Environmental
Operating Humidity
93% RH @ 35 °C (non-condensing)
Operating Temperature
-4 to 122 °F (-20 to 50 °C)
Storage Temperature
14 to 95 °F (-10 to 35 °C)
Physical Features
Dimensions (L x W x H)
5.7 x 2.2 x 1.9 in (14.5 x 5.6 x 4.8 cm)
Weight
≤0.88 lb (0.4 kg)
Enclosure & Protection
Plastic injection with rubber overmold; protection rating IP67 according to IEC 60529
Power
Battery Specs
Internal single cell Li-ion; operational battery life ≥36h; replaceable back-up CR123A with ≥18h operational battery life; recharge time ≤6h when using AC or USB power source
Input Voltage
100-240 VAC (wall and car adapters and USB cable supplied)
Start Up Time
< 2 mins from cold start
Sampling & Analysis
Sample Introduction
Absorption of EM gamma emissions
Sampling & Analysis
From a few seconds to minutes
Standards Compliance
ANSI N42.32 PRD standard fully compliant, ANSI N42.48 SPRD standard fully compliant, including nuclide identification
Threats
Detects gamma radiation emitted from natural occurrences in the environment, special nuclear material, industrial, or medical material
System Interface
Communication
USB 2.0; mini-B socket; Bluetooth® ≤10m range (removable)
Software
On-board webserver software
Display and Alerts
Memory LCD black and white display
Technology
Dose Rate & Accuracy (Cs-137)
≤100 nSv/h – 250 µSv/h (≤10µRem/h – 25 mRem/h); ±20%
Energy Range [Gamma]
25 keV - 3 MeV
Gamma [CsI]
(18mm)³ Cesium Iodide (CsI) with Silicon Photomultiplier (SiPM)
Gamma Spectrum
1,024 channels; 3 MeV
Service Interval
Recommended five year factory maintenance interval; annual dose rate calibration interval
Technology
Spectroscopic personal radiation detector (SPRD)
Typical Resolution
≤7.5 % FWHM at 662 keV